Detect gas quality upsets before they damage your process. Using Tiger Optics’ HALO H2 hydrogen analyzer, you can verify H2impurity levels with part-per-billion accuracy, drift-free stability and instantaneous response. The HALO H2 is easy and fast to install and effortless to maintain, with built-in zero verification. Its robust design—free of moving parts—results in an analyzer that has a high Mean Time Between Failure (MTBF) rate and a very low Cost of Ownership (CoO).
With its patented catalytic conversion technique, utilizing a minute amount of oxygen to cleanly and safely convert hydrogen to moisture, the HALO H2 offers a fully laser-based solution for continuous quality control of your process.
Detection and Matrix | Range | LDL (3σ) | Precision (1σ) @ zero |
---|---|---|---|
H2 in N2 | 0 – 500 ppm | 8 ppb | 3 ppb |
H2 in Ar | 0 – 200 ppm | 6 ppb | 2.0 ppb |
H2 in He | 0 – 125 ppm | 4 ppb | 1.5 ppb |
H2 in O2 | 0 – 250 ppm | 50 ppb (dry-down limited) |
1.5 ppb |
H2 in Clean Dry Air (CDA) | 0 – 450 ppm | 20 ppb (dry-down limited) |
2.5 ppb |
Performance | |
---|---|
Operating range | See Detection Capability table |
Detection limit (LDL, 3σ/24h) | See Detection Capability table |
Precision (1σ, greater of) | ± 0.75% or 1/3 of LDL |
Accuracy (greater of) | ± 4% or LDL |
Speed of response | < 3 minutes to 95% |
Environmental conditions | 10°C to 40°C, 30% to 80% RH (non-condensing) |
Storage temperature | -10°C to 50°C |
Gas Handling System and Conditions | |
Wetted materials | 316L stainless steel, 10 Ra surface finish |
Leak tested to | 1 x 10-9 mbar l / sec |
Gas connections | 1/4” male VCR inlet and outlet |
Sample inlet pressure | 10 − 125 psig (1.7 − 9.6 bara) |
Sample flow rate | 0.5 slpm (± 20%) |
Sample gases | Most inert matrices |
Gas temperature | Up to 60°C |
Utility gas supply* | Mixture of 1% O2, 99% N2 with ~15 sccm, 20 − 125 psig |
Dimensions, H x W x D | |
Standard sensor | 8.73″ x 19.0″ x 23.6″ (222 mm x 483 mm x 599 mm) |
Weight | |
Standard sensor | 45 lbs (20.4 kg) |
Electrical | |
Alarm indicators | 2 user programmable, 1 system fault, Form C relays |
Power requirements | 90 − 240 VAC, 50/60 Hz |
Power consumption | 200 Watts max. |
Signal output | Isolated 4−20 mA |
User interfaces | 5.7” LCD touchscreen, 10/100 Base-T Ethernet, RS-232 |
802.11g Wireless (optional), Modbus TCP (optional) | |
Patents | |
U.S. Patent #7,277,177 • U.S. Patent # 7,255,836 | |
*O2/N2 supply (maximum 10 ppm H2O and H2 impurity) is required for sample conditioning via catalytic conversion. |
Hi there, I'm Bronwyn. I've been with TES for 15 years and I'm the Category Manager for our Gasmet and CEMS equipment sales. My usual work days are Monday - Wednesday. If there's something I can help you with, please give me a call on 0404 161 353 or send me a message and myself or your local sales support will be in touch shortly!