The COMTEC® 6000 GasEx is an Ex protected analyzer for safe measurement of oxygen and combustibles in gas hazardous environments (gas explosion zones 1 and 2). The explosion-protected gas analyzer COMTEC® 6000 GasEx reacts immediately to process gas changes and is ideal for safe and exact measurements of oxygen and COe in refineries or similar areas with explosion hazards (Gas Ex zones 1 / 2).
The ATEX certifications (II 2G Ex db IIC T3 Gb), successfully completed by a German testing institute, apply to all analyzer components, including those directly exposed to the flue gas. This certification creates confidence in plant safety and is coupled with the convincing quality of ENOTEC products, which consistently withstand the challenges of aggressive process conditions.
Applications
- Refineries
- Ex hazard installations
- Petrochemical plants
- Gas by-product/Mixed gas furnaces
- Natural gas liquefaction plants.
Features
- InSitu measurement in real-time
- Overview of calibration history
- ATEX certification
- Low-maintenance design
- Self-monitoring
- Sensor life expectancy on display
- Gas tight sensor construction.
Technical Data
| Process gas temperature | max. 500 °C – max. 1400 °C with cooling protection tube |
| Measuring principle O₂ | Zirconium oxide|MXP |
| Immersion depth/material | 403 mm to 863 mm (others on request) / Stainless steel (SS316TI) |
| Measuring principle COe | MXP (mixed potential) |
| Ambient temperature | Probe: -40°C to +70°C | Electronics: -20°C to +55°C |
| O₂ – Measuring ranges | Two freely-selectable measuring ranges: 0-2 Vol.% to 0-21 Vol.% (others on request) |
| COe – Measuring ranges | 0 to 5000 ppm |
| Housing | Probe: IP66 | Electronics: IP66 |
| Type of protection | II 2G Ex db IIC T3 Gb |