Skip to Content

MesaLabs Defender Series

MesaLabs Defender Series are the industry standard for accurate, reliable and portable Primary calibrators 3 models for 3 flow standards: • 1% Accuracy Volumetric • Low: 5-500ccm • Medium: 50-5,000ccm • High: 300-30,000ccm

Defender 510 and Defender 520

  • Volumetric accuracy: ± 1% of reading
  • Three flow ranges: Low: 5 – 500 ccm; Medium: 50 – 50,000 ccm; High: 300 – 30,000 ccm
  • Defender 510 volumetric primary flow standard
  • Defender 520 enhanced volumetric primary flow standard
  • Documents temperature and pressure for enhanced audit trail.

Defender 530+

  • Three flow ranges: Low: 5 – 500 ccm; Medium: 50 – 50,000 ccm; High: 300 – 30,000 ccm
  • Volumetric accuracy: ± 0.75% of reading
  • Primary flow standard
  • Standardised accuracy: ± 1% of reading
  • Swagelok fittings

Technical

  • Approximate Time per Reading: 1-15 seconds
  • Reading Styles (with averaging function userselectable from 1 to 100 measurement): Single (manual), Continuous or Burst
  • Gas Compatibility: Non-corrosive, non-condensing, less than 70% humidity
  • Flow Modes: Pressure or Suction
  • Measurement Cell Style: Integrated
  • AC Adapter/Charger: 12V DC, >250ma, 2.5 mm center positive
  • Battery System: 6V rechargeable, sealed lead-acid, 6-8 hrs typical operation
  • Battery Operational Time (5 cycles/min): 3 hrs backlight on, 8 hrs backlight off
  • Pressure and Suction Fittings: 1/4” ID Swagelok® compression fittings, 3/8” fittings on High Model
  • Warranty: 1 year; battery 6 months
  • Storage Temperature: 0-70°C
  • Ambient Humidity: 0-70%, non-condensing
  • Operating Pressure (Absolute): Atmospheric, 15 PSI Max
  • Display: Backlit graphical LCD
  • Data Port (for PC interface): Serial (RS-232)
  • Data Cable (for PC interface): 1 meter (Defender Data Port to PC included)
  • Protective Case: Pelican case available
  • Dimensions (H x W x D): 5.5” x 6” x 3” / 140 x 150 x 75 mm
  • Weight: 29 oz / 820 g
ModelDefender 510Defender 520Defender 530+
 Flow RangesLow: 5-500 ccm

Medium: 50-5,000 ccm

High: 300-30,000 ccm

Low: 5-500 ccm

Medium: 50-5,000 ccm

High: 300-30,000 ccm

Low: 5-500 ccm

Medium: 50-5,000 ccm

High: 300-30,000 ccm

Volumetric Accuracy1%1%0.75%
Standardized AccuracyN/AN/A1%
User Selectable Volumetric UnitsmL/min, L/min, cc/min, cf/minmL/min, L/min, cc/min, cf/minmL/min, L/min, cc/min, cf/min
User Selectable Standardized UnitsN/AN/AsmL/min, sL/min, scc/min, scf/ min
User Selectable Temperature UnitsN/A°C, °F°C, °F
User Selectable Pressure UnitsN/AmmHg, PSI, kPammHg, PSI, kPa
Pressure AccuracyN/A± 3.5 mmHg (typical) ±7.0 mm (max)± 3.5 mmHg (typical) ±7.0 mm (max)
Temperature AccuracyN/A± 0.8 °C (typical) ± 1.3 °C (max)± 0.8 °C (typical) ± 1.3 °C (max)
Industry ApplicationsIndustrial Hygiene, Air QualityIndustrial Hygiene, Air QualityManufacturing, Air Quality, Personal Gas Sampling
Pressure and Suction Fittings1/4″ ID Hose Barb1/4″ ID Hose Barb1/4″ ID Swagelok ® compression fittings, 3/8″ fittings on High Model